  • Rugged, waterproof and easy to use, the HI 9829 is the ideal meters for field measurements of lakes, rivers and seas. It displays 1 to 12 parameters simultaneously from up to 15 user selectable parameters. Combined with one of the HI 76x9829 series probes, the HI 9829 can measure water quality parameters such as pH, ORP, conductivity, turbidity, temperature, ions ions ammonium, nitrate, chloride (NH4+, NO3-—N or Cl-), dissolved oxygen concentration, resistivity, TDS, salinity, and seawater specific gravity. Atmospheric pressure is measured for DO compensation.

    The HI 9829 features a graphic, backlit LCD that scales digits to fit up to 12 parameters and allows full configuration of each parameter measured along with an on-screen graphing capability.

    Other Features:

    • Field replaceable ISO 7027 compliant turbidity sensor
    • Ammonium, chloride and nitrate ISE’s
    • Logging from probe or meter
    • Fully customizable instrument, probe, sensors and measurement specifications
    • Display from 1 to 12 parameters with font dimension adjustment
    • Field replaceable sensors
    • pH/ORP or pH, four electrode EC and galvanic DO sensors
    • Auto-recognition of all sensors
    • Rugged probe with stainless steel tip has a diameter under 2” for wells and pipes
    • Fast Tracker™—Tag I.D. System simplifies periodic monitoring
    • Features a built-in barometer for DO compensation
    • Quick or independent sensor calibration feature
    • Measurement check eliminates erroneous readings
    • Logged data can be displayed as graphs
    • Graphic LCD with backlight
    • USB for PC connectivity
    • Good Laboratory Practice feature with last five calibrations recorded
    • Meter accepts both alkaline and rechargeable batteries
    • Waterproof protection for meter (IP67) and probes (IP 68)
  • Model No HI 9829
    pH Range 0.00 to 14.00 pH/±600.0 mV
    pH Resolution 0.01 pH / 0.1 mV
    pH Accuracy ±0.02 pH / ±0.5 mV
    ORP Range ±2000.0 mV
    ORP Resolution 0.1 mV
    ORP Accuracy ±1.0 mV
    Ammonium-
    Nitrogen Range
    0.02 to 200 ppm (as N)
    Ammonium-
    Nitrogen Resolution
    0.01 ppm to 1 ppm; 0.1 ppm to 200 ppm
    Ammonium-
    Nitrogen Accuracy
    ±5% of reading or 2 ppm, whichever is greater
    Chloride Range 0.6 to 200 ppm
    Chloride Resolution 0.01 ppm to 1 ppm; 0.1 ppm to 200 ppm
    Chloride Accuracy ±5% of reading or 2 ppm, whichever is greater
    Nitrate-Nitrogen Range 0.62 to 200 ppm (as N)
    Nitrate-Nitrogen Resolution 0.01 ppm to 1 ppm; 0.1 ppm to 200 ppm
    Nitrate-Nitrogen Accuracy ±5% of reading or 2 ppm, whichever is greater
    Conductivity Range 0 to 200 mS/cm (absolute EC up to 400 mS/cm)
    Conductivity Resolution manual: 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm; automatic: 1 µS/cm from 0 to 9999 µS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm;
    Conductivity Accuracy ±1% of reading or ±1 µS/cm, whichever is greater
    TDS Range 0 to 400000 mg/L or ppm
    TDS Resolution manual: 1 mg/L (ppm); 0.001 g/L (ppt); 0.01g/L (ppt); 0.1 g/L (ppt); 1 g/L (ppt); autorange scales: 1 mg/L (ppm) from 0 to 9999 mg/L (ppm); 0.01 g/L (ppt) from 10.00 to 99.99 g/L (ppt); 0.1 g/L (ppt) from 100.0 to 400.0 g/L (ppt);
    TDS Accuracy ±1% of reading or ±1 mg/L
    Resistivity Range 0 to 999999 Ω•cm; 0 to 1000.0 kΩ•cm; 0 to 1.0000 MΩ•cm
    Resistivity Resolution dependent on resistivity reading
    Resistivity Accuracy dependent on resistivity reading
    Salinity Range 0.00 to 70.00 PSU
    Salinity Resolution 0.01 PSU
    Salinity Accuracy ±2% of reading or ±0.01 PSU
    Seawater σ Range 0 to 50.0 st, s0, s15
    Seawater σ Resolution 0.1 σt, σ0, σ15
    Seawater σ Accuracy ±1 σt, σ0, σ15
    Turbidity Range 0.0 to 99.9 FNU; 100 to 1000 FNU
    Turbidity Resolution 0.1 FNU from 0.0 to 99.9 FNU; 1 FNU from 100 to 1000 FNU
    Turbidity Accuracy ±0.3 FNU or ±2% of reading
    Dissolved Oxygen Range 0.0 to 500.0%; 0.00 to 50.00 ppm
    Dissolved Oxygen Resolution 0.1%; 0.01 ppm
    Dissolved Oxygen Accuracy greater; 300.0 to 500.0%: ±3% of reading; 0.00 to 30.00 ppm: ±1.5% of reading or 0.10 ppm, whichever is greater; 30.00 ppm to 50.00 ppm: ±3% of reading
    Atm. Pressure Range 450 to 850 mm Hg; 17.72 to 33.46 in Hg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.00 to 113.32 kPa
    Atm. Pressure Resolution 0.1 mm Hg; 0.01 in Hg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa
    Atm. Pressure Accuracy ±3 mm Hg within ±15°C from the temperature during calibration
    Temp. Range -5.00 to 55.00°C; 23.00 to 131.00°F; 268.15 to 328.15K
    Temp. Pressure Resolution 0.01°C; 0.01°F; 0.01K
    Temp. Pressure Accuracy ±0.15°C; ±0.27°F; ±0.15K
    Memory 44,000 records
    Output USB PC Connector
    Output USB PC Connector

     

  • If there is one thing that Hanna Instrument's Multiparameter meter guarantee's is robustness, call it the caterpillar of portable meters. It is one of the most robust field meters that are currently available in the market that measures more than a dozen parameters and many simultaneously. Accompanied with Hanna's decades of experience in developing meters for laboratory and field, the HI 9829 provides lab meter like accuracy in the field.